Introduction:  Endoscopic ultrasound (EUS)-guided drainage of symptomatic pancreatic fluid collections (PFCs) using the Hot-Axios device has recently been associated with a significant risk of bleeding. This adverse event (AE) seems to occur less frequently with the use of a different device, the Spaxus stent. The aim of the current study was to compare the rates of bleeding between the two stents. Methods:  Patients admitted for treatment of PFCs by EUS plus lumen-apposing metal stent in 18 endoscopy referral centers between 10 July 2019 and 28 February 2022 were identified and their outcomes compared using a propensity-matching analysis. Results:  363 patients were evaluated. After a 1-to-1 propensity score match, 264 patients were selected (132 per group). The technical and clinical success rates were comparable between the two groups. Significantly more bleeding requiring transfusion and/or intervention occurred in the Hot-Axios group than in the Spaxus group (6.8% vs. 1.5%; P = 0.03); stent type was a significant predictor of bleeding in both univariate and multivariate regression analyses (P = 0.03 and 0.04, respectively). Bleeding necessitating arterial embolization did not however differ significantly between the two groups (3.0% vs. 0%; P = 0.12). In addition, the Hot-Axios was associated with a significantly higher rate of overall AEs compared with the Spaxus stent (9.8% vs. 3.0%; P = 0.04). Conclusion:  Our study showed that, in patients with PFCs, bleeding requiring transfusion and/or intervention occurred significantly more frequently with use of the Hot-Axios stent than with the Spaxus stent, although this was not the case for bleeding requiring embolization.

Lumen-apposing metal stents for the treatment of pancreatic and peripancreatic fluid collections and bleeding risk: a propensity matched study

Facciorusso, Antonio
2024-01-01

Abstract

Introduction:  Endoscopic ultrasound (EUS)-guided drainage of symptomatic pancreatic fluid collections (PFCs) using the Hot-Axios device has recently been associated with a significant risk of bleeding. This adverse event (AE) seems to occur less frequently with the use of a different device, the Spaxus stent. The aim of the current study was to compare the rates of bleeding between the two stents. Methods:  Patients admitted for treatment of PFCs by EUS plus lumen-apposing metal stent in 18 endoscopy referral centers between 10 July 2019 and 28 February 2022 were identified and their outcomes compared using a propensity-matching analysis. Results:  363 patients were evaluated. After a 1-to-1 propensity score match, 264 patients were selected (132 per group). The technical and clinical success rates were comparable between the two groups. Significantly more bleeding requiring transfusion and/or intervention occurred in the Hot-Axios group than in the Spaxus group (6.8% vs. 1.5%; P = 0.03); stent type was a significant predictor of bleeding in both univariate and multivariate regression analyses (P = 0.03 and 0.04, respectively). Bleeding necessitating arterial embolization did not however differ significantly between the two groups (3.0% vs. 0%; P = 0.12). In addition, the Hot-Axios was associated with a significantly higher rate of overall AEs compared with the Spaxus stent (9.8% vs. 3.0%; P = 0.04). Conclusion:  Our study showed that, in patients with PFCs, bleeding requiring transfusion and/or intervention occurred significantly more frequently with use of the Hot-Axios stent than with the Spaxus stent, although this was not the case for bleeding requiring embolization.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11369/446240
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact