Sterilization processes guarantee the sterility of dental instruments but can negatively affect instrument features by altering their physical and mechanical properties. The endodontic instrumentation can undergo a series of alterations, ranging from corrosion to variation in the cutting angle and then changes in the torsional properties and torsional fatigue resistance. This systematic literature review and meta-analysis aims to investigate alterations to the cutting efficiency of endodontic instruments that are induced by procedures for their disinfection and sterilization. Methodologies adopted for this investigation follow the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analysis) guidelines. The following were used as search terms on PubMed and Scopus: “endodontic sterilization”, “endodontic autoclave”, “cyclic fatigue”, “torsional”, “cutting efficiency”, “sterilization”, “surface characteristics”, and “corrosion”. At the end of the selection process, 36 articles were identified, and seven of them are included in this systematic review. The results of a meta-analysis conducted for the use of 10 autoclaving cycles shows a standardized mean difference (SMD) of 0.80 with a p-value equal to 0.04 with respect to effect on cutting efficiency. The network meta-analysis, through direct and indirect comparison between the different autoclave cycles (0, 1, 5, 10, and 15 cycles), revealed that treatment involving 15 autoclave cycles produced the most robust results in terms of having the greatest effects in terms of altered cutting efficiency with a probability of 57.7% and a SUCRA (surface under the cumulative ranking) of 80%. The alterations in the effects on cutting efficiency appear to be triggered after five cycles of sterilization by heat (autoclave). In conclusion, the meta-analysis of the data indicates that the autoclave sterilization protocol must not be repeated more than five times to preserve cutting efficiency. Within the limitations of this review, we can therefore establish that sterilization by autoclaving alone results in steel and NiTi instruments becoming less efficient in cutting after five cycles, as measured by a reduction in cutting efficiency.

The effects of sterilization procedures on the cutting efficiency of endodontic instruments: A systematic review and network meta-analysis

Dioguardi M.;Di Cosola M.;Laino L.;Troiano G.;Muzio L. L.
2021-01-01

Abstract

Sterilization processes guarantee the sterility of dental instruments but can negatively affect instrument features by altering their physical and mechanical properties. The endodontic instrumentation can undergo a series of alterations, ranging from corrosion to variation in the cutting angle and then changes in the torsional properties and torsional fatigue resistance. This systematic literature review and meta-analysis aims to investigate alterations to the cutting efficiency of endodontic instruments that are induced by procedures for their disinfection and sterilization. Methodologies adopted for this investigation follow the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analysis) guidelines. The following were used as search terms on PubMed and Scopus: “endodontic sterilization”, “endodontic autoclave”, “cyclic fatigue”, “torsional”, “cutting efficiency”, “sterilization”, “surface characteristics”, and “corrosion”. At the end of the selection process, 36 articles were identified, and seven of them are included in this systematic review. The results of a meta-analysis conducted for the use of 10 autoclaving cycles shows a standardized mean difference (SMD) of 0.80 with a p-value equal to 0.04 with respect to effect on cutting efficiency. The network meta-analysis, through direct and indirect comparison between the different autoclave cycles (0, 1, 5, 10, and 15 cycles), revealed that treatment involving 15 autoclave cycles produced the most robust results in terms of having the greatest effects in terms of altered cutting efficiency with a probability of 57.7% and a SUCRA (surface under the cumulative ranking) of 80%. The alterations in the effects on cutting efficiency appear to be triggered after five cycles of sterilization by heat (autoclave). In conclusion, the meta-analysis of the data indicates that the autoclave sterilization protocol must not be repeated more than five times to preserve cutting efficiency. Within the limitations of this review, we can therefore establish that sterilization by autoclaving alone results in steel and NiTi instruments becoming less efficient in cutting after five cycles, as measured by a reduction in cutting efficiency.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11369/403635
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 9
social impact