Objectives The aim of this study was to evaluate the incidence and clinical implications of leaks (acute incomplete occlusion, early and late reopenings) following LAA ligation with the LARIAT device. Background Percutaneous LAA ligation with the LARIAT device may represent an alternative for stroke prevention in high-risk patients with atrial fibrillation with contraindications to oral anticoagulation. Methods This was a retrospective, multicenter study of 98 consecutive patients undergoing successful LAA ligation with the LARIAT device. Leaks were defined as the presence of flow as evaluated by transesophageal echocardiography (TEE). TEE was performed during the procedure, at 6 and 12 months, and after thromboembolic events. Results Leaks were detected in 5 (5%), 14 (15%), and 19 (20%) patients at the 3 time points. During follow-up, 5 patients developed neurological events (4 strokes and 1 transient ischemic attack). Two occurred early (1 fatal stroke and 1 stroke with multiple recurrences in the following months), and TEE was not repeated after the events. The remaining 3 occurred late (after 6 months) and were associated with small leaks (<5 mm). In 2 of 3 cases, such a small leak was missed by the standard evaluation on 2-dimensional TEE, being evident only with the aid of 3-dimensional imaging. Conclusions Incomplete occlusion of the LAA after LARIAT ligation is relatively common and may be associated with thromboembolic events. Proper long-term surveillance with careful TEE should be considered to detect leaks, which can be managed with either resumption of oral anticoagulation or percutaneous transcatheter closure.

Clinical Implications of Leaks Following Left Atrial Appendage Ligation with the LARIAT Device

DI BIASE, LUIGI;
2016-01-01

Abstract

Objectives The aim of this study was to evaluate the incidence and clinical implications of leaks (acute incomplete occlusion, early and late reopenings) following LAA ligation with the LARIAT device. Background Percutaneous LAA ligation with the LARIAT device may represent an alternative for stroke prevention in high-risk patients with atrial fibrillation with contraindications to oral anticoagulation. Methods This was a retrospective, multicenter study of 98 consecutive patients undergoing successful LAA ligation with the LARIAT device. Leaks were defined as the presence of flow as evaluated by transesophageal echocardiography (TEE). TEE was performed during the procedure, at 6 and 12 months, and after thromboembolic events. Results Leaks were detected in 5 (5%), 14 (15%), and 19 (20%) patients at the 3 time points. During follow-up, 5 patients developed neurological events (4 strokes and 1 transient ischemic attack). Two occurred early (1 fatal stroke and 1 stroke with multiple recurrences in the following months), and TEE was not repeated after the events. The remaining 3 occurred late (after 6 months) and were associated with small leaks (<5 mm). In 2 of 3 cases, such a small leak was missed by the standard evaluation on 2-dimensional TEE, being evident only with the aid of 3-dimensional imaging. Conclusions Incomplete occlusion of the LAA after LARIAT ligation is relatively common and may be associated with thromboembolic events. Proper long-term surveillance with careful TEE should be considered to detect leaks, which can be managed with either resumption of oral anticoagulation or percutaneous transcatheter closure.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11369/340405
Citazioni
  • ???jsp.display-item.citation.pmc??? 12
  • Scopus 43
  • ???jsp.display-item.citation.isi??? ND
social impact